AED Servicing in Ontario: Battery, Pad and Inspection Dates

Nov 12
9 min read

A defibrillator on the wall is not a rescue plan. It is a piece of equipment with a shelf life, and the uncomfortable truth is that public AEDs can fail before the emergency, not during it. Expired pads with dried gel, batteries quietly past their service window, and missing inspection records are the usual culprits. None of these failures announce themselves. They wait.


This tutorial sets out a named servicing schedule for facilities and venues across Ontario, with dates, owners and documentation built in. You will work through why public AEDs fail before the rescue; what servicing actually means for a public AED; the monthly inspection routine and readiness lights; battery replacement intervals, prompts and the label trap; pad expiry and why gel dries out first; the log that proves readiness; fault signals and what to do; the Ontario context and what the rules do and do not say; when to bring in a service provider; quick answers; and your next thirty days.


By the end, you will have a workable schedule and a clear owner for every task. That is what separates a wall-mounted sign from a device that will actually work.


Why Public AEDs Fail Before the Rescue, Not During It


Most AED failures trace to upkeep, not the rescue. Allina Health maintenance guidance identifies a dead battery as the most common cause of malfunction, and a unit can sit on the wall looking ready while holding no usable charge. Pads are sealed around a conductive gel that can dry out, and manufacturer guidance cites pad life of 2 to 4 years, sometimes shorter than the battery interval. Replacing the battery alone does not restore readiness.


Any warning is local to the device: a red light or beep at the cabinet, with no guarantee of a building alarm, a fire panel signal, or notice to the person who walks past each morning. Without install dates, expiry labels and a monthly check log, a facility manager cannot show when the device was last verified.


Treat each wall-mounted AED as a consumable device, not a permanent fixture. Batteries, pads and records have finite lives; a dated AED maintenance schedule with a named owner is what separates a wall sign from working equipment.


What Servicing Actually Means for a Public AED


Servicing a public AED is not one task but three, and a program that handles only one of them is incomplete.


The physical side is presence and placement: is the unit where the map says it is, is the cabinet or carry case intact, is the location unobstructed and signed, is the readiness indicator showing ready, and is the unit free of audible alarms?


The consumable side covers pads and battery: are the sealed pads in date, is the spare set in date, and is the battery within its replacement window or being prompted by the device? Replace by the manufacturer interval or when the unit asks, whichever comes first.


Documentation is what most sites skip: install dates, pad and battery expiry dates, monthly initialled check entries, fault reports, and a record of any service work completed on the unit.


There is no single universal schedule. Build your written schedule from your own device manual, then confirm it against your workplace emergency response plan.


The Inspection Schedule: Monthly Checks and Readiness Lights


Monthly visual checks are the baseline for a typical AED; workplace emergency response plans may require more frequent checks, and high-traffic venues may choose to run checks more frequently, weekly is a defensible option, though no source mandates it.


Monthly checks confirm nine things:


  • Present, correctly located

  • Cabinet/case undamaged

  • Readiness indicator reads ready

  • No beeping or alarm tone

  • Pads pouch sealed and in date

  • Spare pads in date

  • Battery within interval

  • Accessory kit complete

  • Signage legible


Readiness Lights and Event-Driven Checks


Readiness indicators are model-specific; commonly green means ready, red with 30-second beeping means fault. Record what yours mean and post it inside the cabinet door.


Add event-driven checks after use, accidental activation during cleaning or moving, pad or battery swap, and building work or relocation that could disturb the cabinet. Book an annual program review beyond the device: audit the log for gaps, confirm the named owner is still in role, verify current expiry dates, and consider confirming staff hold current CPR-C/AED certification as part of an annual program review. This discipline makes regular AED maintenance a system, not a one-off task.


AED Battery Replacement: Intervals, Prompts and the Label Trap


Eventually, the readiness check will find a battery at the end of its service life. Manufacturer guidance cited by Allina Health puts battery life in the 2 to 5 year range, tied to the model rather than one industry standard. Confirm the interval printed in your own manual before adding it to a schedule.


The label trap catches careful people. The date printed on an AED battery is usually a use-by or shelf-life date, not necessarily the expiry once installed. Write the install date, or the calculated expiry, on the device at the moment of replacement.


Low-battery prompts vary by model, so a warning is a deadline to act, never something to defer to the next scheduled check. As a practical rule of thumb, no specific lead time is set by regulation, ordering 60 to 90 days before expiry allows time for delivery. Battery life figures also assume storage as the manufacturer specifies, so check your location against the manual's temperature guidance and keep the unit in its cabinet rather than an unheated room.


AED Pad Expiry: Why Gel Dries Out Before the Battery Dies


Battery replacement is only half of consumable readiness. AED pads are single-use electrodes coated in a conductive gel that can dry out over time, and pads may expire on a shorter timeline than the battery, the manufacturer's stated range is 2 to 4 years. Allina Health guidance cites several common models on 2-year cycles.


Replace both the attached set and the spare set. Spares bought with the device usually share the same expiry, and an expired spare is not a spare. Check the pouch as carefully as the printed date: punctures, creases, water stains or swelling can compromise gel before expiry. Replace damaged packaging.


If you stock adult and paediatric electrodes, track them separately. They are distinct products with distinct expiry dates, storage positions and reorder needs. Record pad install and expiry dates alongside battery details; the shorter date should drive reminders. Replace pads after any patient use or opened training pouch. Treat demonstration pads as single-use and do not return them to rescue-ready stock.


Documentation and Ownership: The Log That Proves Readiness


Getting the dates right is only half the job; the other half is proving them. As best practice (no regulation prescribes this), name one person, not a department, plus a named deputy alongside them. That individual owns the monthly check, the reorder reminder and the record itself, which is accountability a shared facilities inbox never produces.


A workable AED log carries these fields:


  • Device location and model, serial number, installation date

  • Battery install date and expiry

  • Pad install date and expiry, plus spare pad expiry

  • Monthly check date with initials

  • Faults observed with the action taken

  • Service calls with the work completed


Keep it in two places: a log card with the device, findable by a responder or inspector in seconds, and a digital copy that survives staff turnover, fire-warden changes and building handovers.


Every entry should show a next due date, so the log answers what is coming and who is responsible. Consider keeping current CPR-C/AED certifications alongside the log as part of a complete program. Apply the same rigour here as to credentials and compliance documentation: these records support insurance conversations, post-incident reviews and internal audits.


What to Do When Your AED Signals a Fault


Your log becomes the response tool when the indicator changes. Follow the same sequence every time.


Confirm the signal before acting. On many models a steady green light means ready, while a red light with beeping at 30-second intervals signals a fault. Check your manual's indicator legend, as symbols and tones vary by manufacturer.


Then run the simple checks:


  • Confirm the pads are seated correctly and in date.

  • Confirm the battery is properly seated and within its replacement window.

  • Confirm the cabinet has not been disturbed.


The most common culprits, dead battery or expired pads, are covered in the Battery and Pad sections above; confirm both before escalating.


If the fault persists, take the unit out of service or mark it clearly as unserviceable. An AED left mounted and faulty can be pulled for a real emergency and fail.


Log the fault at once: date, indicator state, checks performed, action taken. Book the service or replacement, then record when the device returned to service so the coverage gap is documented.


Plan for downtime before it happens. High-footfall or remote sites should identify a temporary alternative or escalation route in advance.


After service work, confirm the readiness indicator is normal before signing the unit back into the log as rescue-ready.


Ontario Context: What the Rules Do and Do Not Say


No single Ontario provincial standard dictates AED inspection intervals, registration or record-keeping. Ontario's Defibrillator Registration and Public Access Act, 2020 was passed but, as of current reporting, remains unproclaimed with no regulations published. Confirm any Health Canada or Ontario obligation directly before relying on it.



In the absence of a confirmed provincial schedule, practical obligations flow from three places: the manufacturer's owner manual for your specific model, which governs replacement intervals; your workplace first aid and emergency response program, which governs check frequency; and your organisation's own safety planning, alongside related compliance work such as Onsite Respirator Fit Testing in Ontario: A Compliance Guide for Employers.


That variability is why documentation matters more than any single number. Whatever the regulatory position, a log showing what was checked, when, and by whom is the practical standard a facility can defend.


Public access programs that place more AEDs in community settings do not transfer maintenance obligation away from the device owner. A donated or grant-funded unit carries the same battery, pad and documentation requirements as one purchased outright.


Where a site spans multiple buildings, registrations, insurers or tenants, standardise on one schedule and one log format across all devices so nothing is lost at the boundary between responsible parties.


When to Bring In an AED Service Provider


Where the rules leave the decision to your own program, capacity becomes the practical test. Bring in an AED service provider when devices span multiple buildings, no one can reliably run the monthly check, several expiry dates converge, a fault will not clear, or documentation has lapsed.


Ask any provider to state visit scope in writing: checks performed, consumables replaced, condition the unit is left in, and the record you receive. Confirm that scope against your exact models before the first visit.


Forest City First Aid offers AED kit servicing alongside first aid kit servicing; confirm current AED servicing scope and coverage area on its service page. Pair it with training: CPR-C/AED, CPR-HCP/BLS, Intermediate First Aid, Airway Management and Naloxone Training give a maintained device a certified responder.


Enter all replacements and dates into your existing log (see Documentation section).


Use the first visit as a baseline audit. Request a device-by-device summary of model, serial number, battery date, pad date and faults, then build the next twelve months of in-house checks from it.


Quick Answers on AED Servicing and Inspection


Inspection: Monthly visual checks are the baseline minimum. Workplace emergency response plans may require more; high-traffic venues often choose to. The owner's manual governs specifics.


Pads: Usually 2 to 4 years by manufacturer, with several common models on 2-year cycles. Replace attached and spare sets by packaging date, and any pouch with a damaged seal regardless of date.


Battery life: Guidance cites 2 to 5 years, model-specific. (See Battery section above for the label trap.)


Dead battery: Check the readiness indicator and listen for prompts or tones. Treat a low-battery warning as urgent.


Ontario registration: This article does not confirm a provincial requirement. Verify any Health Canada or Ontario obligation directly, and treat the manual, employer first aid program and safety plan as working requirements.


Your Next Thirty Days of AED Servicing


With those answers settled, convert them into a dated plan. Pull the manual for every AED on site and record its battery interval, pad interval and inspection frequency in one schedule, with a named owner for each device.


Run a first pass today: verify each readiness indicator, note printed expiry dates on all batteries and pad sets including spares, and replace anything already out of date first.


Open the log with serial numbers, install and expiry dates, and the first monthly check entry. Keep a copy with the device and a digital copy, and set reorder reminders well before each expiry date.


Post the indicator legend inside each cabinet, brief staff on fault signals and who to notify, then review the schedule annually with your CPR-C/AED certification records.


Conclusion


Your public AED is a life-saving device only if it is ready to use. The key takeaways are clear: batteries and pads expire on different schedules, monthly readiness checks catch problems early, and proper documentation proves your device is rescue-ready. Your next step is simple: check your AED today. Verify the readiness indicator, note all expiry dates, and update your log. This single action starts a habit of readiness that protects everyone who counts on your device. A well-maintained AED is more than equipment; it is a promise to your community. Make that promise real, one check at a time.

 
 
 